GUI Application Message List
| Message | Message Details | How to Resolve |
|---|---|---|
| A file necessary for startup is lacked. Please reinstall control software. | Required startup files are missing. Reinstall the control software. | Reinstall the software because required runtime files are missing. |
| Buffer Over Run Serial : 13XXXXX - X State : XXX Please re-connect all "MicroPeckerX" and refresh device. | Buffer overrun. Serial: 13XXXXX - X State: XXX Reconnect all MicroPeckerX units and refresh devices. | Contact the developer/vendor. |
| Configuration Load Error. Error : XXX | Configuration load error. Error: XXX | Load a correct configuration file. |
| Configuration Load Error. File not exists. | Configuration load error. File does not exist. | Check file name/path of the configuration file. |
| Configuration Load Error. File version is Illegal. | Configuration load error. File version is invalid. | Load a configuration file with correct format/version. |
| Configuration Load Error. Folder is not exists. | Configuration load error. Folder does not exist. | Check folder path. |
| Configuration Load Error. Illegal file format. | Configuration load error. Invalid file format. | Load a correctly formatted configuration file. |
| Configuration Save Error. Folder is not exists. | Configuration save error. Folder does not exist. | Check destination folder path. |
| Currently logging. Are you sure you want to abort? | Logging is currently running. Abort? | If aborted, saved logs remain, but consistency across channels/time is not guaranteed. |
| Currently under monitoring. Are you sure you want to terminate? | Monitoring is running. Stop monitoring? | - |
| Delete layout "XXX"? | Delete layout XXX? | - |
| Error of generating log replay buffer. | Failed to generate log replay buffer. | Ensure sufficient disk free space. |
| Failed to ready for log replay thread. | Failed to prepare log replay thread. | PC may be low on resources. Use a higher-spec PC. |
| Failed to ready for monitor thread. | Failed to prepare monitor thread. | PC may be low on resources. Use a higher-spec PC. |
| Failed to save the log. | Failed to save log. | Check whether destination file is open in another application. |
| Internal Error Serial : 13XXXXX - X Error Code : XXX State : XXX Please re-connect all "MicroPeckerX" and refresh device. | Internal error. Serial: 13XXXXX - X Error Code: XXX State: XXX Reconnect all MicroPeckerX units and refresh devices. | Contact the developer/vendor. |
| It failed save error log. | Failed to save error log. | Check whether destination file is open in another application. |
| Layout "XXX" already exists. Overwrite? | Layout XXX already exists. Overwrite? | - |
| Logging canceled. The contents of the log file are not guaranteed. | Logging canceled. Log file content is not guaranteed. | If logging is interrupted, content consistency cannot be guaranteed. |
| MicroPeckerX Communication Error Serial : 13XXXXX - X Error Code : XXX Please re-connect all "MicroPeckerX" and refresh device. | MicroPeckerX communication error. Serial: 13XXXXX - X Error Code: XXX Reconnect all MicroPeckerX units and refresh devices. | Contact the developer/vendor. |
| "MicroPeckerX Control Software" is already started. | MicroPeckerX Control Software is already running. | Multiple instances are not allowed. |
| MicroPeckerX information acquisition failed. | Failed to acquire MicroPeckerX information. | Reinstall because required files may be missing. |
| Monitor synchronization error. Check the connection of the synchronous cable. | Monitor synchronization error. Check sync cable connection. | When using multiple units, connect all units with sync cable. |
| No setting information of the connected device. Display the import dialog? | No setting information for connected device. Open import dialog? | If loaded settings do not include connected device info, execute import. |
| Set layout "XXX" as default? | Set layout XXX as default? | - |
| Setting Error. [Simulation] The file of log replay is not load. Serial : 13XXXXX Ch : X | Setting error. [Simulation] Log replay file is not loaded. Serial: 13XXXXX Ch: X | Load the file before starting monitor. |
| Setting Error. [Trigger] In the ReTriggered mode, must set the start trigger. | Setting error. [Trigger] Start trigger is required in ReTriggered mode. | Enable/configure start trigger. |
| Setting Error. The key trigger is duplicated. Key:XXX ChX-X XXX ChX-X XXX | Setting error. Duplicate key trigger. Key: XXX ChX-X XXX ChX-X XXX | Check duplicated triggers and assign unique keys. |
| Setting Error. The log folder does not exist. | Setting error. Log folder does not exist. | Check log save folder path. |
| The Help file is not found. | Help file not found. | Reinstall software. |
| The log folder number has reached its maximum. | Log folder number reached maximum. | Select another destination folder. |
| There are no logging data. | No logging data. | Check cable connection between MicroPeckerX and target. |
| Setting is cleared. Are you sure? | Settings will be cleared. Continue? | - |
| Delete the display filter? | Delete display filter? | - |
| Delete the label filter "XXX"? | Delete label filter XXX? | - |
| Load Error. | Load error. | Check whether Hex/Binary file format is correctly described. |
| Event Trigger is not selected. Are you sure to continue? | Event trigger is not selected. Continue? | If event transmission mode is selected, configure event trigger. |
| Log File Header Error. | Log file header error. | Load a correctly formatted log file. |
| Log file is not found. | Log file not found. | Check file path and file name. |
| Log File Read Error. Line:XXX | Log file read error. Line: XXX | Load a correctly formatted log file. |
| The number of maximum transmission logs is exceeded. | Maximum number of transmission logs exceeded. | Log replay / burst send supports up to 1.2 million records. Extra records are ignored. |
| Are you sure to clear list? | Clear the list? | - |
| Failed to save. | Failed to save. | Check whether destination file is open in another application. |
| Log File Incompatible Error. | Log file compatibility error. | In Analyze tab, MicroPecker-format logs cannot be loaded directly. Load in Ch Log tab first, then run mixing. |
| Old Format Log Convert Result Chx-x : Valid Log = XXX, Invalid Log = YYY | Old format log conversion result. Chx-x: Valid log = XXX, Invalid log = YYY | Unsupported logs for MicroPeckerX are discarded. Review invalid logs before use. |
| The same Ch can not be specified. | The same channel cannot be specified. | Set different source and destination channels for gateway. |
| There are no valid analytical logs. | No valid analysis logs. | May occur when filtering removes all logs. Review filter settings. |