GUI Application Message List
| Message | Description | Corrective Action |
|---|---|---|
| A file necessary for startup is lacked. Please reinstall control software. | A required startup file is missing. Please reinstall the control software. | Reinstall the software because required runtime files are missing. |
| Buffer Over Run Serial : 12XXXXX - X State : XXX Please re-connect all "MicroPeckerX" and refresh device. | Buffer overrun detected. Serial : 12XXXXX - X State : XXX Reconnect all MicroPeckerX devices and refresh devices. | Contact support. |
| Configuration Load Error. Error : XXX | Configuration load error. Error : XXX | Load the correct configuration file. |
| Configuration Load Error. File not exists. | Configuration load error. File not found. | Check that the configuration file name is correct. |
| Configuration Load Error. File version is Illegal. | Configuration load error. Invalid file version. | Load a configuration file in the correct format. |
| Configuration Load Error. Folder is not exists. | Configuration load error. Folder not found. | Check that the configuration file path is correct. |
| Configuration Load Error. Illegal file format. | Configuration load error. Invalid file format. | Load a configuration file in the correct format. |
| Configuration Save Error. Folder is not exists. | Configuration save error. Folder not found. | Check that the save destination path is correct. |
| Currently logging. Are you sure you want to abort? | Logging is in progress. Abort logging? | If logging is aborted, saved logs remain, but consistency across channels is not guaranteed. |
| Currently under monitoring. Are you sure you want to terminate? | Monitoring is in progress. Stop monitoring? | - |
| Delete layout "XXX"? | Delete layout "XXX"? | - |
| Error of generating log replay buffer. | Failed to generate log replay buffer. | Ensure sufficient free disk space. |
| Failed to ready for log replay thread. | Failed to initialize log replay thread. | Remaining disk/memory resources may be insufficient. Use a higher-spec PC. |
| Failed to ready for monitor thread. | Failed to initialize monitor thread. | Remaining disk/memory resources may be insufficient. Use a higher-spec PC. |
| Failed to save the log. | Failed to save log. | Check whether the destination file is open in another application. |
| Internal Error Serial : 12XXXXX - X Error Code : XXX State : XXX Please re-connect all "MicroPeckerX" and refresh device. | Internal error. Serial : 12XXXXX - X Error Code : XXX State : XXX Reconnect all MicroPeckerX devices and refresh devices. | Contact support. |
| It failed save error log. | Failed to save error log. | Check whether the destination file is open in another application. |
| Layout "XXX" already exists. Overwrite? | Layout "XXX" already exists. Overwrite? | - |
| Logging canceled. The contents of the log file are not guaranteed. | Logging was canceled. Log file contents are not guaranteed. | If logging is aborted, saved logs remain, but consistency across channels is not guaranteed. |
| MicroPeckerX Communication Error Serial : 12XXXXX - X Error Code : XXX Please re-connect all "MicroPeckerX" and refresh device. | MicroPeckerX communication error. Serial : 12XXXXX - X Error Code : XXX Reconnect all MicroPeckerX devices and refresh devices. | Contact support. |
| "MicroPeckerX Control Software" is already started. | MicroPeckerX Control Software is already running. | Multiple instances are not supported. |
| MicroPeckerX information acquisition failed. | Failed to acquire MicroPeckerX information. | Reinstall software because required files may be missing. |
| Monitor synchronization error. Check the connection of the synchronous cable. | Monitor synchronization error. Check sync cable connection. | When using multiple units, connect MicroPeckerX units with sync cables. |
| No setting information of the connected device. Display the import dialog? | No setting information for the connected device. Open import dialog? | If connected device information is missing in loaded settings, run import processing. |
| Set layout "XXX" as default? | Set layout "XXX" as default? | - |
| Setting Error. [Device] Ack must be enabled when use simulation mode. Serial : 12XXXXX Ch : X | Setting error. [Device] Ack must be enabled when simulation mode is used. Serial : 12XXXXX Ch : X | Enable Ack for channels set to simulation/log replay/burst mode. |
| Setting Error. [Simulation] Log replay mode or burst transmission mode can not be used at the same time both Ch. Serial : 12XXXXX Ch : X | Setting error. [Simulation] Log replay mode and burst transmission mode cannot be used on both channels simultaneously. Serial : 12XXXXX Ch : X | For each MicroPeckerX, set log replay or burst mode on only one channel. |
| Setting Error. [Simulation] The file of burst transmit is not load. Serial : 12XXXXX Ch : X | Setting error. [Simulation] Burst transmit file is not loaded. Serial : 12XXXXX Ch : X | Load the file before starting monitoring. |
| Setting Error. [Simulation] The file of log replay is not load. Serial : 12XXXXX Ch : X | Setting error. [Simulation] Log replay file is not loaded. Serial : 12XXXXX Ch : X | Load the file before starting monitoring. |
| Setting Error. [Trigger] In the ReTriggered mode, must set the start trigger. | Setting error. [Trigger] Start trigger must be set in ReTriggered mode. | Enable start trigger in ReTriggered mode. |
| Setting Error. The key trigger is duplicated. Key:XXX ChX-X XXX ChX-X XXX | Setting error. Duplicate key trigger. Key: XXX ChX-X XXX ChX-X XXX | Check duplicate triggers and avoid assigning the same key. |
| Setting Error. The log folder does not exist. | Setting error. Log folder does not exist. | Check that the log save path is correct. |
| The Help file is not found. | Help file not found. | Reinstall software. |
| The log folder number has reached its maximum. | Log folder number has reached the maximum. | Select another destination folder. |
| There are no logging data. | No logging data. | Check cable connections; device-target connection may be incorrect. |
| Setting is cleared. Are you sure? | Settings will be cleared. Continue? | - |
| Delete the display filter? | Delete display filter? | - |
| Delete the label filter "XXX"? | Delete label filter "XXX"? | - |
| [CAN/CAN-FD Data Setting Dialog] Load Error. | Load error. | Verify Hex/Binary format file contents are correct. |
| Delete the selected slot? | Delete selected slot? | - |
| Event Trigger is not selected. Are you sure to continue? | Event trigger is not selected. Continue? | If Event transmission mode is selected, configure an event trigger. |
| Log File Header Error. | Log file header error. | Load a log in the correct format. |
| Log file is not found. | Log file not found. | Check file path. |
| Log File Read Error. Line:XXX | Log file read error. Line: XXX | Load a log in the correct format. |
| The number of maximum transmission logs is exceeded. | Maximum number of transmittable logs exceeded. | Log replay/burst supports up to 1.2 million records; excess records are ignored. |
| Are you sure delete the frame information? | Delete frame information? | - |
| Are you sure delete the gateway information? | Delete gateway information? | - |
| Are you sure to clear list? | Clear list? | - |
| Failed to save. | Failed to save. | Check whether the destination file is open in another application. |
| It failed in analyzing of frame. | Frame analysis failed. | Run analysis with at least one valid (checked) frame setting. |
| It failed in analyzing of gateway. | Gateway analysis failed. | Run analysis with at least one valid (checked) gateway setting. |
| Log File Incompatible Error. | Log file compatibility error. | In Analyze tab, MicroPecker format logs cannot be loaded directly; load in Ch tab first, then execute Mix. |
| Old Format Log Convert Result Chx-x : Valid Log = XXX, Invalid Log = YYY | Old format log conversion result. Chx-x : Valid Log = XXX, Invalid Log = YYY | Logs unsupported by MicroPeckerX are discarded. Check if invalid logs cause issues. |
| The same Ch can not be specified. | The same channel cannot be specified. | Set different channels for gateway source and destination. |
| There are no valid analytical logs. | No valid analysis logs. | This appears when logs are fully filtered out. Review filter settings. |