MessageDescriptionCorrective Action
A file necessary for startup is lacked.
Please reinstall control software.
A required startup file is missing. Please reinstall the control software.Reinstall the software because required runtime files are missing.
Buffer Over Run
Serial : 12XXXXX - X
State : XXX
Please re-connect all "MicroPeckerX" and refresh device.
Buffer overrun detected.
Serial : 12XXXXX - X
State : XXX
Reconnect all MicroPeckerX devices and refresh devices.
Contact support.
Configuration Load Error.
Error : XXX
Configuration load error.
Error : XXX
Load the correct configuration file.
Configuration Load Error.
File not exists.
Configuration load error.
File not found.
Check that the configuration file name is correct.
Configuration Load Error.
File version is Illegal.
Configuration load error.
Invalid file version.
Load a configuration file in the correct format.
Configuration Load Error.
Folder is not exists.
Configuration load error.
Folder not found.
Check that the configuration file path is correct.
Configuration Load Error.
Illegal file format.
Configuration load error.
Invalid file format.
Load a configuration file in the correct format.
Configuration Save Error.
Folder is not exists.
Configuration save error.
Folder not found.
Check that the save destination path is correct.
Currently logging.
Are you sure you want to abort?
Logging is in progress.
Abort logging?
If logging is aborted, saved logs remain, but consistency across channels is not guaranteed.
Currently under monitoring.
Are you sure you want to terminate?
Monitoring is in progress.
Stop monitoring?
-
Delete layout "XXX"?Delete layout "XXX"?-
Error of generating log replay buffer.Failed to generate log replay buffer.Ensure sufficient free disk space.
Failed to ready for log replay thread.Failed to initialize log replay thread.Remaining disk/memory resources may be insufficient. Use a higher-spec PC.
Failed to ready for monitor thread.Failed to initialize monitor thread.Remaining disk/memory resources may be insufficient. Use a higher-spec PC.
Failed to save the log.Failed to save log.Check whether the destination file is open in another application.
Internal Error
Serial : 12XXXXX - X
Error Code : XXX
State : XXX
Please re-connect all "MicroPeckerX" and refresh device.
Internal error.
Serial : 12XXXXX - X
Error Code : XXX
State : XXX
Reconnect all MicroPeckerX devices and refresh devices.
Contact support.
It failed save error log.Failed to save error log.Check whether the destination file is open in another application.
Layout "XXX" already exists. Overwrite?Layout "XXX" already exists. Overwrite?-
Logging canceled.
The contents of the log file are not guaranteed.
Logging was canceled.
Log file contents are not guaranteed.
If logging is aborted, saved logs remain, but consistency across channels is not guaranteed.
MicroPeckerX Communication Error
Serial : 12XXXXX - X
Error Code : XXX
Please re-connect all "MicroPeckerX" and refresh device.
MicroPeckerX communication error.
Serial : 12XXXXX - X
Error Code : XXX
Reconnect all MicroPeckerX devices and refresh devices.
Contact support.
"MicroPeckerX Control Software" is already started.MicroPeckerX Control Software is already running.Multiple instances are not supported.
MicroPeckerX information acquisition failed.Failed to acquire MicroPeckerX information.Reinstall software because required files may be missing.
Monitor synchronization error.
Check the connection of the synchronous cable.
Monitor synchronization error.
Check sync cable connection.
When using multiple units, connect MicroPeckerX units with sync cables.
No setting information of the connected device.
Display the import dialog?
No setting information for the connected device.
Open import dialog?
If connected device information is missing in loaded settings, run import processing.
Set layout "XXX" as default?Set layout "XXX" as default?-
Setting Error.
[Device] Ack must be enabled when use simulation mode.
Serial : 12XXXXX
Ch : X
Setting error.
[Device] Ack must be enabled when simulation mode is used.
Serial : 12XXXXX
Ch : X
Enable Ack for channels set to simulation/log replay/burst mode.
Setting Error.
[Simulation] Log replay mode or burst transmission mode can not be used at the same time both Ch.
Serial : 12XXXXX
Ch : X
Setting error.
[Simulation] Log replay mode and burst transmission mode cannot be used on both channels simultaneously.
Serial : 12XXXXX
Ch : X
For each MicroPeckerX, set log replay or burst mode on only one channel.
Setting Error.
[Simulation] The file of burst transmit is not load.
Serial : 12XXXXX
Ch : X
Setting error.
[Simulation] Burst transmit file is not loaded.
Serial : 12XXXXX
Ch : X
Load the file before starting monitoring.
Setting Error.
[Simulation] The file of log replay is not load.
Serial : 12XXXXX
Ch : X
Setting error.
[Simulation] Log replay file is not loaded.
Serial : 12XXXXX
Ch : X
Load the file before starting monitoring.
Setting Error.
[Trigger] In the ReTriggered mode, must set the start trigger.
Setting error.
[Trigger] Start trigger must be set in ReTriggered mode.
Enable start trigger in ReTriggered mode.
Setting Error.
The key trigger is duplicated.
Key:XXX
ChX-X XXX
ChX-X XXX
Setting error.
Duplicate key trigger.
Key: XXX
ChX-X XXX
ChX-X XXX
Check duplicate triggers and avoid assigning the same key.
Setting Error.
The log folder does not exist.
Setting error.
Log folder does not exist.
Check that the log save path is correct.
The Help file is not found.Help file not found.Reinstall software.
The log folder number has reached its maximum.Log folder number has reached the maximum.Select another destination folder.
There are no logging data.No logging data.Check cable connections; device-target connection may be incorrect.
Setting is cleared.
Are you sure?
Settings will be cleared.
Continue?
-
Delete the display filter?Delete display filter?-
Delete the label filter "XXX"?Delete label filter "XXX"?-
[CAN/CAN-FD Data Setting Dialog]
Load Error.
Load error.Verify Hex/Binary format file contents are correct.
Delete the selected slot?Delete selected slot?-
Event Trigger is not selected. Are you sure to continue?Event trigger is not selected. Continue?If Event transmission mode is selected, configure an event trigger.
Log File Header Error.Log file header error.Load a log in the correct format.
Log file is not found.Log file not found.Check file path.
Log File Read Error. Line:XXXLog file read error. Line: XXXLoad a log in the correct format.
The number of maximum transmission logs is exceeded.Maximum number of transmittable logs exceeded.Log replay/burst supports up to 1.2 million records; excess records are ignored.
Are you sure delete the frame information?Delete frame information?-
Are you sure delete the gateway information?Delete gateway information?-
Are you sure to clear list?Clear list?-
Failed to save.Failed to save.Check whether the destination file is open in another application.
It failed in analyzing of frame.Frame analysis failed.Run analysis with at least one valid (checked) frame setting.
It failed in analyzing of gateway.Gateway analysis failed.Run analysis with at least one valid (checked) gateway setting.
Log File Incompatible Error.Log file compatibility error.In Analyze tab, MicroPecker format logs cannot be loaded directly; load in Ch tab first, then execute Mix.
Old Format Log Convert Result
Chx-x : Valid Log = XXX, Invalid Log = YYY
Old format log conversion result.
Chx-x : Valid Log = XXX, Invalid Log = YYY
Logs unsupported by MicroPeckerX are discarded. Check if invalid logs cause issues.
The same Ch can not be specified.The same channel cannot be specified.Set different channels for gateway source and destination.
There are no valid analytical logs.No valid analysis logs.This appears when logs are fully filtered out. Review filter settings.