GUI Application Message List

MessageMessage ContentCorrective Action
A file necessary for startup is lacked.
Please reinstall control software.
Required startup file is missing. Reinstall control software.Reinstall the software because required files are missing.
Buffer Over Run
Serial : 13XXXXX - X
State : XXX
Please re-connect all "MicroPeckerX" and refresh device.
Buffer overrun.
Serial : 13XXXXX - X
State : XXX
Reconnect all MicroPeckerX units and refresh devices.
Contact the developer/support.
Configuration Load Error.
Error : XXX
Configuration load error.
Error: XXX
Load a valid configuration file.
Configuration Load Error.
File not exists.
Configuration load error.
File does not exist.
Check that the configuration filename is correct.
Configuration Load Error.
File version is Illegal.
Configuration load error.
Invalid file version.
Load a configuration file in the correct format.
Configuration Load Error.
Folder is not exists.
Configuration load error.
Folder does not exist.
Check that the configuration file path is correct.
Configuration Load Error.
Illegal file format.
Configuration load error.
Invalid file format.
Load a configuration file in the correct format.
Configuration Save Error.
Folder is not exists.
Configuration save error.
Folder does not exist.
Check that the save destination path is correct.
Currently logging.
Are you sure you want to abort?
Logging is in progress.
Abort?
If logging is aborted, saved logs remain, but consistency across all channels is not guaranteed.
Currently under monitoring.
Are you sure you want to terminate?
Monitoring is in progress.
Stop monitoring?
-
Delete layout "XXX"?Delete layout "XXX"?-
Error of generating log replay buffer.Failed to generate log replay buffer.Ensure sufficient free disk space.
Failed to ready for log replay thread.Failed to prepare log replay thread.PC resources (disk/memory) may be insufficient. Use a higher-spec PC.
Failed to ready for monitor thread.Failed to prepare monitor thread.PC resources (disk/memory) may be insufficient. Use a higher-spec PC.
Failed to save the log.Failed to save log.Check whether destination file is open in another application.
Internal Error
Serial : 13XXXXX - X
Error Code : XXX
State : XXX
Please re-connect all "MicroPeckerX" and refresh device.
Internal error.
Serial : 13XXXXX - X
Error Code : XXX
State : XXX
Reconnect all MicroPeckerX units and refresh devices.
Contact the developer/support.
It failed save error log.Failed to save error log.Check whether destination file is open in another application.
Layout "XXX" already exists. Overwrite?Layout "XXX" already exists. Overwrite?-
Logging canceled.
The contents of the log file are not guaranteed.
Logging canceled.
Log file content is not guaranteed.
If logging is aborted, saved logs remain, but consistency across all channels is not guaranteed.
MicroPeckerX Communication Error
Serial : 13XXXXX - X
Error Code : XXX
Please re-connect all "MicroPeckerX" and refresh device.
MicroPeckerX communication error.
Serial : 13XXXXX - X
Error Code : XXX
Reconnect all MicroPeckerX units and refresh devices.
Contact the developer/support.
”MicroPeckerX Control Software” is already started.MicroPeckerX Control Software is already running.Multiple instances are not supported.
MicroPeckerX information acquisition failed.Failed to acquire MicroPeckerX information.Reinstall because required files may be missing.
Monitor synchronization error.
Check the connection of the synchronous cable.
Monitor synchronization error.
Check sync cable connection.
When using multiple units, connect units with sync cable.
No setting information of the connected device.
Display the import dialog?
No settings for connected device.
Open import dialog?
If connected device info is missing in loaded settings, run import.
Set layout "XXX" as default?Set layout "XXX" as default?-
Setting Error.
[Simulation] The file of log replay is not load.
Serial : 13XXXXX
Ch : X
Setting error.
[Simulation] Log replay file is not loaded.
Serial : 13XXXXX
Ch : X
Load file before starting monitoring.
Setting Error.
[Trigger] In the ReTriggered mode, must set the start trigger.
Setting error.
[Trigger] Start trigger must be set in ReTriggered mode.
Enable start trigger in ReTriggered mode.
Setting Error.
The key trigger is duplicated.
Key:XXX
ChX-X XXX
ChX-X XXX
Setting error.
Duplicate key trigger.
Key: XXX
ChX-X XXX
ChX-X XXX
Check duplicated triggers and avoid setting the same key.
Setting Error.
The log folder does not exist.
Setting error.
Log folder does not exist.
Check that log save path is correct.
The Help file is not found.Help file was not found.Reinstall software.
The log folder number has reached its maximum.Log folder number reached maximum.Select another save destination folder.
There are no logging data.No logging data.Check cable connections between MicroPeckerX and target.
Setting is cleared.
Are you sure?
Settings will be cleared.
Continue?
-
Delete the display filter?Delete display filter?-
Delete the label filter "XXX"?Delete label filter "XXX"?-
Load Error.Load error.Check whether Hex/Binary format file content is correctly described.
Event Trigger is not selected. Are you sure to continue?Event trigger is not selected. Continue?If event transmission mode is selected, configure event trigger.
Log File Header Error.Log file header error.Load a log file in correct format.
Log file is not found.Log file not found.Check file path, etc.
Log File Read Error. Line:XXXLog file read error. Line: XXXLoad a log file in correct format.
The number of maximum transmission logs is exceeded.Maximum transmission log count exceeded.Up to 1.2 million records can be loaded for log replay/log burst transmission. Excess records are discarded.
Are you sure to clear list?Clear the list?-
Failed to save.Failed to save.Check whether destination file is open in another application.
Log File Incompatible Error.Log file compatibility error.In Analyze tab of Analyze Window, MicroPecker format logs cannot be loaded directly. Load once in Ch Log tab, then run Mix.
Old Format Log Convert Result
Chx-x : Valid Log = XXX, Invalid Log = YYY
Old-format log conversion result.
Chx-x : Valid Log = XXX, Invalid Log = YYY
Logs for MicroPecker that are unsupported by MicroPeckerX are discarded. Check invalid logs before use.
The same Ch can not be specified.Same channel cannot be specified.Set different channels for gateway source and destination.
There are no valid analytical logs.No valid analysis logs.Displayed when filtering removes all logs. Review filter settings.